PDD-1600 is probably the most powerful diode driver in the aesthetic world. Peak output power is up to 10kW (with user selectable IMAX and VMAX). Averaged output power is up to 1600W by default (assuming 100-240VAC input) and over 2000W on request (with 200-240VAC input).
Since the driver was especially designed for laser hair removal applications, it's fully compliant to the medical standards - IEC 60601-1 for electrical safety and IEC 60601-1-2 for electromagnetic compatibility.
As a result PDD-1600 can be unattendedly used in medical devices.
Output:
Maximum average power
1600 W (2000 W on request)
Maximum current (
I
max
)
selectable up to 200 A
Maximum voltage (
U
max
)
selectable up to 100 V
Maximum peak power (
W
max
)
10 kW (15kW on request)
Rise / fall time
< 1 ms (10% to 90% full current),
< 500 μs on request
Current regulation accuracy
< 1% of
I
max
Current value error
< 1% of
I
max
Current overshoot
< 1% of
I
max
Interface:
Connector
15 Pin "D"-Sub Female
Current program
analog, 0-10V
Environment:
Operating temperature
0 … +40 °C
Storage temperature
-20 … +60 °C
Humidity
90%, non-condensing
Other:
Size (LxWxH)
230x223x133 mm (see also the dimensional drawing)